3.5 Basic Operations
3-54 ZM2376
When you execute both of the OPEN and SHORT corrections, or when you execute the LOAD
correction, any of OPEN, SHORT, or LOAD correction is not executed and the measured value
becomes CORR Err unless the following condition is met.
Condition for correction : OPEN correction value (|Z|) > SHORT correction value (|Z|) × 2
When the OPEN and SHORT correction values are close, the error will increase unless the LOAD
correction is executed. If the ratio is 1000 times or less, it is recommended to execute the LOAD
correction.