3.5 Basic Operations
3-61 ZM2376
Setting numeric values/reading of spot SHORT correction value
You can set the spot SHORT correction value at the current measurement frequency with a numeric
value or can read the value to fine tune it.
In the spot SHORT correction menu, first specify the format and then set the value.
_SHIFT_
+ [SHORT], _3_ key operations >> Spot SHORT correction menu
SPOT Correction: OFF Format:Ls-Rs @SHORT
0) OFF 1) ON 2) Format 3) Entry
Format >> Format setting menu for spot SHORT correction value
Format: Ls-Rs @SHORT Current setting
0) Rs-X 1) Ls-Rs Options
Rs Residual resistance (unit: Ω) Rs X Rs Ls
X Residual reactance (unit: Ω)
Ls Residual inductance (unit: H)
If a setting is made or the EXIT operation is performed, one-previous menu comes back.
Entry >> Spot SHORT correction value input menu
Example:Correction value for the Rs-X format
Co-SHORT=Rs: +123.456mΩ X : +123.456mΩ Current value
±(0.0000p to 999999M) Z<900Ω 1.00000kHz Available ranges
Spot SHORT correction frequency
Frequency when the correction value is measured or set
If disabled, "---"
Set the measured value in the short state measured with all of the OPEN, SHORT, and LOAD
corrections disabled. Initially, the cursor exists on the previous value. When the previous value is
input and entered with the _ENTR_
key, the cursor moves to the next value, indicating that the
next value can be input. The cursor can be moved between previous and next values by pressing
the _BS_ | [PREV] or _EXP_ | [NEXT] key. If either one of the values is set, the measurement
frequency at that time is recorded as the spot SHORT correction frequency.
Perform the EXIT operation to return to one-previous menu.
If the spot SHORT correction value is set, the SHORT correction is disabled. To use the set
correction value, enable both the spot correction and the SHORT correction.
Though the warning message will be displayed for the input of correction value of impedance ≥
900Ω, it is set as it is as the spot SHORT correction value.
If the spot SHORT correction value is zero, the measured value remains as it was before the
correction even when the correction is executed.