EasyManua.ls Logo

Nortec UL60 - I-Charge Functionality

Nortec UL60
119 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
9 Tabular overview of the program procedures I-Charge
104 © Nortec Electronics 2018
9.6 I-Charge
With the help of program “P5: I-Charge” individual charging programs can be started.
Charging time span and final charging voltage free selection possible.
Example:
NiCd battery
Enter:
1.55 V shut down voltage
maximum 1:00 hour charging time span
20: NiCd battery (manufacturer independent)
P520: PARAMETER INPUT
MAX. CHARGE TIME 01:00 H
20: NiCd battery (manufacturer independent)
P520: PARAMETER INPUT
MAX. VOLTAGE/CELL ? 1.55 V/C
================================
COPYRIGHT
NORTEC
2014-15 UL60 V 1.0 16.06.15
SN: 160001
--------------------------------
DATE: 28.06.15
BATTERY NO:
OPERATOR NO:
NOM. VOLTAGE: 24.00 V
RATED CAP: 040.0 Ah
PROGRAM: 520
NiCd battery
I-Charge
--------------------------------
PROGRAM START 10:38:02
VOLTAGE = 25.41 V
--------------------------------
P520:S1: 11:35:53
I-CHARGE
VOLTAGE = 31.00 V
CURRENT = 40.01 A
CAPACITY = 38.07 Ah
= 096 %
TIME = 057 MIN
--------------------------------
PROGRAM AUTOMATICALLY FINISHED
================================
================================
COPYRIGHT
NORTEC
2014-15 UL60 V 1.0 16.06.15
SN: 160001
--------------------------------
DATE: 28.06.15
BATTERY NO:
OPERATOR NO:
NOM. VOLTAGE: 24.00 V
RATED CAP: 040.0 Ah
PROGRAM: 520
NiCd battery
I-Charge
--------------------------------
PROGRAM START 10:38:02
VOLTAGE = 25.41 V
--------------------------------
P520:S1: 11:38:03
I-CHARGE
VOLTAGE = 30.55 V
CURRENT = 40.01 A
CAPACITY = 40.00 Ah
= 100 %
TIME = 060 MIN
--------------------------------
PROGRAM AUTOMATICALLY FINISHED
================================
Termination because shut down
voltage has been reached.
Termination because maximum
charging time span has been
reached.

Table of Contents