9 Tabular overview of the program procedures Discharge
© Nortec Electronics 2018 55
9.1.5 Program 019: Discharge Pb battery (free input)
19:Pb battery (free input)
select Batteryprocessing:
P0: Discharge
Enter:
Nominal voltage
Rated capacity
Discharging current C1, C2, C5, C10 or C20
Discharging final voltage of each cell (selectable 1.5 - 1.9 V)
Step 1:
Discharging until the voltage of final discharging voltage for each cell is reached.
End of program
Time monitoring:
Discharging maximum 24 hours.
9.1.6 Program 020: Discharge NiCd battery (manufacturer independent)
20: NiCd battery (manufacturer independent)
select Batteryprocessing:
P0: Discharge
Enter :
Number of cells
Rated capacity
Cut-off voltage per cell
Step 1:
Discharging with C
1
until a voltage of 1 V for each cell is reached.
In case no automatc cell tester is connected:
After 48 minutes a signal asks for the measurement of the single cells voltages.
Otherwise: cell measurement takes place automatically.
End of program
Example:
Connected is a 12 V battery with 40 Ah.
The battery will be discharged with 40 A to 10 V.
Time monitoring:
Discharging maximum 90 minutes.