6
INSTALLATION
hhmmss TIME Set current time to hh: mm: ss, hh is between 0 - 23.
TIME Print current time.
mmddyy DATE Set current date mmddyy.
DATE Print current date.
% Print current % ref.
# Print current ref., any function.
AC Abort calibration /Test.
C Start an auto calibration.
E Go to initial state of current function.
F Print current function.
M Same as mode button.
xD Set 1 second print delay (set x = 0 for OFF, or x = 1 for ON).
ID Print current ID string.
xxxxID Enter ID string (xxxx limit to 8 characters).
RS232 COMMAND TABLE
Field: Mode Stab CR LF
Length: 5 1 1 1
blank if stable
? if unstable
Command
Character Description
? Print current unit.
0 = minimum level
1=
2 = maximum level
xI Set averaging level to x, where x = 0, 1, 2
Grams, Dwt, Carat, Oz Av, Oz t, Grain, Tael 1,
Tael 2, Tael 3, Momme, Custom Unit
xM Places balance in mode x, where x = 1 to 11
(see table).
If unit is not already enabled,
command will be ignored.
1 = grams
2 = pennyweight
3 = carats
4 = avoidupois ounces
5 = troy ounces
6 = grains
7 = taels
8 = mommes
11 = custom unit
P Print display data
When numeric only displaydata is
selected for output in the RS232 menu,
the Mode field is not output.
Field: Weight Mode Stab CR LF
Length: 10 1 5 1 1 1
Same as ?
command
Displayed weight sent right justified
w/lead zero blanking.
Nine characters include:
decimal point (1)
weight (7 max))
polarity (1): blank if positive
- if negative
Menu
Menu