7
INSTALLATION
Command
Character Description
xS Set stable data only printing (set x = 0 for OFF, or x = 1 for ON.
T Same effect as pressing on tare button.
V Print EPROM version
x# Set current ref., any function, CW takes two reference separated by a space.
x% Doanloads reference weight “x” for percent mode. “x” must be in grams. Command is
ignored if percent mode is disabled. If percent mode is enabled, balance will automatically
switch to percent mode display.
xF Set current function to “x”. x = 0 to 6. Setup menu must be unlocked.
0 = None
1 = Percent
2 = Parts Counting
3 = Check Weighing
4 = Animal Weighing
5 = Fill Guide
6 = High Point
xZ Set Auto Zero to “x”. x = 0 for OFF, x = 1 for ON
xSL Set stability level. User menu must be unlocked. x = 0 to 3
Z Zero request (Gross tare) if Net/Gross enabled.
nnnA Set Auto Print feature to “nnn” (see table).
nnn = 0 Turns feature OFF
nnn = S Output on stability
nnn = C Output is continuous
nnn = 1-256 Sets Auto Print interval
Esc L Prints listing of Setup and Print menu settings.
Esc R Resets Setup and Print menus to factory defaults.
CAUTION: This will reset RS232 configuration.
Esc S Save current settings.
Field: Model # 1 EPROM # CR LF
Length: 6 16 1 1
Balance Model
“98101-xx Sr#x.xx”