  • T
    Todd WilcoxAug 31, 2025
    What to do if Orphee Measuring Instruments displays 'BACKUP : NO MEMORY AVAILABLE FOR STORAGE'?
    • P
      patricia44Aug 31, 2025
      If the Orphee Measuring Instruments displays 'BACKUP : NO MEMORY AVAILABLE FOR STORAGE', delete the stored results.
  • J
    joshuajonesSep 9, 2025
    What to do if Orphee Measuring Instruments displays 'CYCLE: DOOR OPENED'?
    • M
      Melinda LangSep 9, 2025
      If the Orphee Measuring Instruments displays 'CYCLE: DOOR OPENED', close the door.
  • V
    victoriajohnsonAug 29, 2025
    What does it mean if Orphee Measuring Instruments displays 'BACKUP : MEMORY IS ALMOST FULL. PLEASE DELETE RESULTS'?
    • L
      Lisa HurleyAug 29, 2025
      If the Orphee Measuring Instruments displays 'BACKUP : MEMORY IS ALMOST FULL. PLEASE DELETE RESULTS', delete results.
  • C
    Curtis DavisSep 23, 2025
    Why is Orphee Measuring Instruments not printing?
    • J
      jameswilliamsSep 23, 2025
      If the Orphee Measuring Instruments displays 'No printing', check the paper and the electrical connection.
  • H
    hayestracySep 10, 2025
    What to do if Orphee MYTHIC 18 Measuring Instruments displays 'INTERN: RESULT AREA IS LOCKED'?
    • B
      Brian ParkerSep 12, 2025
      If the Orphee Measuring Instruments displays 'INTERN: RESULT AREA IS LOCKED', wait before performing a cycle. If the issue persists, restart the Mythic.
  • R
    Rachel SmithSep 14, 2025
    What to do if Orphee Measuring Instruments displays 'THE CLEANER USED IS OUT OF DATE'?
    • J
      Julia GibsonSep 14, 2025
      If the Orphee Measuring Instruments displays 'THE CLEANER USED IS OUT OF DATE', replace the bottle and perform a prime Cleaner.
  • R
    Randall TuckerSep 23, 2025
    Why is diluent leaking around the needle during the run cycle of Orphee Measuring Instruments?
    • V
      vjensenSep 23, 2025
      If the Orphee Measuring Instruments displays 'Diluent leaks around the needle during the run cycle', check the rinsing needle block for clots and clean it if necessary (see section 9.3.2).
  • C
    Cassandra BrownAug 27, 2025
    How to troubleshoot Orphee Measuring Instruments when it's not starting?
    • M
      Matthew McleanAug 27, 2025
      If the Orphee Measuring Instruments is not starting, check the power supply connection wires.
  • P
    ptorresAug 26, 2025
    What to do if all results from Orphee MYTHIC 18 Measuring Instruments are bad?
    • T
      Tammy GilbertAug 26, 2025
      If all results from the Orphee Measuring Instruments are bad, check the diluent level and ensure that the tubing is not pinched.
  • H
    Heather MullinsAug 25, 2025
    What to do if Orphee MYTHIC 18 shows 'No results' because there is no HCT & PLT?
    • D
      David GarnerAug 26, 2025
      If the Orphee Measuring Instruments displays 'No results' because there is no HCT & PLT, check the bath wires. If the issue persists, perform a Cleaning Cycle, and if still unsuccessful, follow with a Bleach cycle.

Summary

IMPORTANT SAFETY INSTRUCTIONS

General Safety Warnings

Key safety warnings including reading manual, operator training, and handling metal edges.

Electromagnetic Immunity Guidance

INTRODUCTION

Installation Safety Reminder

Emphasizes reading the user's manual before installation or maintenance.

FLUIDICS SYSTEM

Fluidics Module Overview

Fluidics Parts Description

ELECTRIC & ELECTRONIC COMPONENTS

Board Overview and ESD Precautions

Lists the three main boards and provides critical ESD handling instructions.

Electronic System Diagram

Wiring Diagram - General View

DATA OUTPUT FORMAT

Routine Results Data Format

PARTS LIST AND BLOW-UP VIEWS

SOFTWARE OPERATION

Parameter Adjustment Procedures

INSTRUMENT MAINTENANCE

Maintenance Safety Precautions

Mandatory safety measures, including wearing gloves and using disinfectant.

Concentrated Cleaning Procedure

Clean Out Procedure for Transportation

REPAIRING AND REPLACEMENT

Repair Safety and Emergency Stop

Emphasizes safety measures and the emergency stop procedure.

Counting Baths Module Repair

Syringe Module Repair

Complete Syringe Module Replacement

Electronic Board Replacement Precautions

Main Board Replacement Procedure

Fuse and Battery Replacement

Preamplifier Board Replacement

TROUBLESHOOTING

Analytical Problem Diagnosis

Table addressing issues with WBC, RBC, and HGB results and their solutions.

Troubleshooting Error Messages

Orphee MYTHIC 18 Specifications

General IconGeneral
BrandOrphee
ModelMYTHIC 18
CategoryMeasuring Instruments
LanguageEnglish