Why is Orphee Measuring Instruments not printing?
- JjameswilliamsSep 23, 2025
If the Orphee Measuring Instruments displays 'No printing', check the paper and the electrical connection.
Why is Orphee Measuring Instruments not printing?
If the Orphee Measuring Instruments displays 'No printing', check the paper and the electrical connection.
Why is Orphee Measuring Instruments printing badly?
If the Orphee Measuring Instruments displays 'Bad printing', check the black ribbon.
What to do if Orphee MYTHIC 18 displays 'BACKUP : FOLDER NOT FOUND'?
If the Orphee Measuring Instruments displays 'BACKUP : FOLDER NOT FOUND', restart the Mythic.
What to do if Orphee MYTHIC 18 Measuring Instruments displays 'CYCLE : INIT NOT DONE'?
If the Orphee Measuring Instruments displays 'CYCLE : INIT NOT DONE', perform an initialization or a control cycle.
What to do if Orphee MYTHIC 18 shows 'No results' and there is no HGB?
If the Orphee Measuring Instruments displays 'No results' and there is no HGB, check the bath wires, the lysis level, and ensure the right lysis tubing is correctly connected to the WBC bath.
What to do if Orphee MYTHIC 18 shows 'No results' because there is no HCT & PLT?
If the Orphee Measuring Instruments displays 'No results' because there is no HCT & PLT, check the bath wires. If the issue persists, perform a Cleaning Cycle, and if still unsuccessful, follow with a Bleach cycle.
What to do if all results from Orphee MYTHIC 18 Measuring Instruments are bad?
If all results from the Orphee Measuring Instruments are bad, check the diluent level and ensure that the tubing is not pinched.
How to troubleshoot Orphee Measuring Instruments when it's not starting?
If the Orphee Measuring Instruments is not starting, check the power supply connection wires.
What to do if Orphee MYTHIC 18 shows bad stability?
If the Orphee Measuring Instruments shows bad stability, perform a Back flush and a Cleaning Cycle. If unsuccessful, then perform a Bleach cycle. Also, check the level bubble flow in the WBC bath during the run cycle and the lysis quantity (see section 7.2.2.2).
What to do if Orphee MYTHIC 18 shows bad stability with HCT & PLT too?
If the Orphee Measuring Instruments shows bad stability with HCT & PLT too, perform a Back flush and a Cleaning Cycle. If unsuccessful, then perform a Bleach cycle. Also, check the level bubble flow in the RBC bath during the run cycle and the level bubble flow in the WBC bath during the first dilution.
Key safety warnings including reading manual, operator training, and handling metal edges.
Emphasizes reading the user's manual before installation or maintenance.
Lists the three main boards and provides critical ESD handling instructions.
Mandatory safety measures, including wearing gloves and using disinfectant.
Emphasizes safety measures and the emergency stop procedure.
Table addressing issues with WBC, RBC, and HGB results and their solutions.
| Brand | Orphee |
|---|---|
| Model | MYTHIC 18 |
| Category | Measuring Instruments |
| Language | English |
Loading...