DigitalFlow™ DF868 Multipurpose Ultrasonic Liquid Flowmeter Programming Manual (2-Channel) 15
Chapter 1. Programming Site Data
1.6.3.2 Pipe Material
4. Use the [F1]-[F4] and [
 keys to select the Pipe Material, as listed in Table 7.
Note: Depending on the pipe material choice, another window may appear, asking you to specify the specific
material. If you have selected “Other,” the DF868 prompts you to enter the sound speed.
1.6.3.3 Pipe OD
5. Enter the known Pipe OD or circumference and press [F1]-[F4] to select the appropriate units, as shown in
Table 8. Press [ENT]. (The meter will only accept values from 1/8 to 648 in.)
Note: The first two lines of text in the prompt area depend on the selection made at the TRANSDUCER NUMBER
prompt. The option bar choices may appear in English or Metric units.
Obtain the required information by measuring either the pipe outside diameter (OD) or circumference at the
transducer installation site. The data may also be obtained from standard pipe size tables.
Table 7: Limits for I/O Parameters
Pipe Material Category Specific Material
Steel Carbon Steel or Stainless Steel
Iron Ductile Iron or Cast Iron
Cu - Copper None
Al - Aluminum None
Brass None
CuNi - Copper/Nickel 70% Cu 30% Ni or 90% Cu 10% Ni
Glass Pyrex, Flint, or Crown
Plastic Nylon, Polyester, Polypropylene, PVC (CPVC), Acrylic
Other* Any material
Table 8: Available Pipe OD Units
English Metric
inch = pipe OD in inches mm = pipe OD in millimeters
feet = pipe OD in feet m = pipe OD in meters
in/PI = pipe circumference in inches mm/PI = pipe circumference in millimeters
ft/PI = pipe circumference in feet m/PI = pipe circumference in meters