DigitalFlow™ DF868 Multipurpose Ultrasonic Liquid Flowmeter Programming Manual (2-Channel) 71
Chapter 3. Logging Data
3.3 Stopping a Log
Use the STOP submenu to terminate a logging process that is currently active. Press the appropriate side of the
[SCREEN] key to activate the desired pane of the display screen, and complete the following steps:
1. To access the Log Menu, press [LOG].
a.
If you have activated the security feature (see Chapter 1, Programming Site Data, page 50 for details), enter the
assigned password and press
[ENT].
Note: Once a log is stopped it cannot be restarted, but the log remains in memory. To clear the log from memory,
refer to Chapter 5, Clearing Data, page 5-4, for instructions.
2. Press [F3] = STOP.
3. Use the [
, [ and [F1]-[F4] keys to select a log file to stop.
Note: At the NAME prompt, the option bar will show the names of all completed or active log files that are
currently stored in memory. Recall that a flashing asterisk (*) appears at the far right of the locator bar if the
Model DF868 is currently logging data (see Chapter 3, Operation, of the Startup Guide).
4. Press [F1] to continue logging and return to the initial Log Menu screen. Press [F2] to stop logging and return to
the initial Log Menu screen.
Procedure Options
Do one of the following:
Press [F1]-[F4] to enter another [LOG] submenu.
Press [EXIT] to return to the data screen.