The maximum permitted weight at the detector’s inlet must be no more than 5 kg and the maxi-
mum torque must be 3 N · m.
The inlet pressure must be no higher than atmospheric pressure. Pressure that is too high can
damage the product.
The detector's performance depends on the type of accessories used and on the quality of the
mechanical connections.
When assembling the vacuum circuit, use accessories to shut off the product and make mainte-
nance easier (inlet shut off valves, purge systems, etc.).
Comply with these recommendations to optimize measurement (see chapter “Pre-requisites for
optimizing use”).
Connection
Remove the blanked-off flange that covers the detector’s inlet port and save it for reuse during
storage or transport.
Use pipes with a diameter equal to the diameter of the detector’s inlet. The pipes should be as
short as possible and completely sealed.
Connect the part or the installation to be tested using the connection accessories available in the
product catalogue.
Connect the part or the installation to be tested using the flexible pipeworks. Never use rigid or
flexible plastic pipework (compressed air style tube).