The maximum permitted weight at the detector’s inlet must be no more than 5 kg and the maxi-
mum torque must be 3 N · m.
●
The inlet pressure must be no higher than atmospheric pressure. Pressure that is too high can
damage the product.
●
The detector's performance depends on the type of accessories used and on the quality of the
mechanical connections.
●
When assembling the vacuum circuit, use accessories to shut off the product and make mainte-
nance easier (inlet shut off valves, purge systems, etc.).
●
Comply with these recommendations to optimize measurement (see chapter “Pre-requisites for
optimizing use”).
Connection
►
Remove the blanked-off flange that covers the detector’s inlet port and save it for reuse during
storage or transport.
►
Use pipes with a diameter equal to the diameter of the detector’s inlet. The pipes should be as
short as possible and completely sealed.
►
Connect the part or the installation to be tested using the connection accessories available in the
product catalogue.
►
Connect the part or the installation to be tested using the flexible pipeworks. Never use rigid or
flexible plastic pipework (compressed air style tube).
Installation
23/98