8.4.6 Burn-in
This function is used to prepare the detector, leaving it in optimal working condition by automatically
carrying out a series of short tests and inlet vents between each test.
NOTICE
Risk of pollution
Before starting this function, make sure that the leak detector is in an environment free of tracer
gas pollution.
Access: Menu [Maintenance] [Burn-in]
Inlet vent Only available if the inlet vent is set to ‘Operator’.
Access the ‘Inlet vent’ menu
From the ‘Inlet vent’ menu, return to the ‘Burn-in’ menu by pressing [<].
Start without calib. Function to be started
Function launching
Series of tests and inlet vents
Start with calibration Function to be started
Test runs, inlet vents and calibrations
Available in hard vacuum test only
Stop Stop burn-in in progress
Prerequisites
Detector in ‘Stand-by’ mode
“Automatic” inlet vent selected
Procedure
1. Implement the preliminary conditions.
2. Install a blank-off flange on the detector’s inlet port.
3. Press [Start without calib.] or [Start with calib.].
4. Press [Stop] or the START/STOP button to stop the burn-in.
8.4.7 Internal Pirani gauge calibration
This function is used to calibrate the detector’s Pirani internal gauge.
Access: Menu [Maintenance] [Internal Pirani calibration]
Pressure Read only
Limit pressure or atmospheric pressure display depending on the procedural step.
Pressure ≈ 5000 -> Limit pressure display
Pressure ≈ 30000 -> Atmospheric pressure display
Inlet pressure Read only
Inlet pressure display.
HV valid Function launching
Setting the limit pressure (≈ 5000)
Hard Vacuum Displaying the limit pressure
Atm valid Function launching
Setting the atmospheric pressure (≈ 30000)
Atmospheric Displaying the atmospheric pressure
Start/Stop Function launching
Test start
Inlet vent Function launching
Prerequisites
Hard vacuum test method selected
Most sensitive test mode selected
Manual cycle end (= 'operator' selected).
Settings menu
71/98