EasyManua.ls Logo

Pfeiffer Vacuum ASM 310 - Calibration; Calibration type

Pfeiffer Vacuum ASM 310
98 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7. Then run the sniffer probe slowly over the areas of the part to be tested that may leak: the leak
rate displayed varies when a leak is detected (quantitative value of the measured leak rate).
8. Stop the test by pressing the START/STOP button.
7.5 Calibration
Calibration is used to ensure that the leak detector is correctly adjusted to detect the tracer gas selected
and display the correct leak rate.
A calibrated leak is used to calibrate the leak detector.
By default, internal leak is selected to allow for a rapid calibration of the leak detector. It is also possible
to calibrate the leak detector using an external leak.
The leak detector should be calibrated with a calibrated leak of the same type as the tracer gas used.
Detector calibration
20 minutes after switching it on, the detector suggests that the user perform a calibration.
For correct use of the detector and to optimize the accuracy of the measurement, this cali-
bration must be performed.
It is recommended to perform a calibration:
at least once a day,
for intense operation: start calibration at the beginning of each work session (e.g. work
in shifts, every 8 hours),
if it is uncertain whether the detector is working properly.
Internal calibrated leak
The internal calibrated leak is specifically designed for the leak detector. It is composed of:
a reservoir filled with
4
He tracer gas (no internal calibration with other tracer gases),
a temperature sensor (to take into account the effect of temperature on the leak rate),
an integrated diaphragm (to calibrate the leak rate),
an identification label (identical to the identification label of an external calibrated leak).
The calibrated leak is supplied with a calibration certificate.
Use a calibrated leak in the range of ≈ 10
-7
mbar · l/s (≈ 10
-8
Pa · m
3
/s).
In the case of intensive use of the detector, a spare internal calibrated leak is recommend-
ed. By default, the detector can be calibrated with an external calibrated leak.
External calibrated leak
The operator must use a calibrated leak containing the tracer gas selected (
4
He, Mass 3 or H
2
).
There are several types of external calibrated leaks, with or without reservoir, with or without valve, cov-
ering several ranges of leaks.
The manufacturer does not provide calibrated leaks in Mass 3 and H
2
.
The choice of external calibrated leak depends on application requirements: use a calibrat-
ed leak from the same leak rate range as the leak to be measured.
7.5.1 Calibration type
The user assigns a calibration type to the CAL button.
Operation
27/98

Table of Contents

Other manuals for Pfeiffer Vacuum ASM 310

Related product manuals