EasyManua.ls Logo

Pfeiffer Vacuum ASM 310 - Information

Pfeiffer Vacuum ASM 310
98 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
1 2 3 4
1 Chronological calibration number 3 Calibration result (see below)
2 Date and time of the calibration 4 Indicator of analyzer cell performance (see below)
Result Description
Success [Cell sta-
tus: xxx %]
Calibration success
[Cell status: xxx %] Indicator of analyzer cell performance.
Default settings: between 90% and 100%
Normal operation: between 10% and 100%
Normal wear on some cell components will reduce this val-
ue over time but will not reduce the accuracy of the detec-
tor's measurements.
FAIL Calibration failure
Cal. check OK Calibration check successful
Cal. check FAIL Calibration check failure
History export
An export can be generated containing the event and calibration history.
2 possible modes of access:
[Maintenance] [History] [Events history]
[Maintenance] [History] [Calibrations history]
1. Insert a USB stick into the control panel.
2. Press [Export].
The message “Events and calibrations exported” is displayed to confirm the export.
8.4.2 Information
This function is used to view information on the leak detector.
Settings menu
67/98

Table of Contents

Other manuals for Pfeiffer Vacuum ASM 310

Related product manuals