21 3 4
1 Chronological event number 3 Code for the event
2 Date and time of the event 4 Description of the event
Information coding:
Code Event Description
I300 Air inlet Inlet vent
I301 Stop on pollution Test stops automatically if measured leak rate pollution > Pollu-
tion
I302 Reset timer backing pump Backing pump timer reset
I303 Reset timer turbo pump 1 Secondary pump 1 hour timer reset
I306 Reset timer filament #1 Filament 1 hour timer reset
I307 Reset timer filament #2 Filament 2 hour timer reset
I308 Reset cycles counter Cycle timer reset
I309 Emission increase
4
He, Mass 3: change in emission intensity (Ie)
2
H: change in emission intensity (Ie)
I310 Calibration restart Automatic start of a new calibration
I311 Detector stop Shutdown the detector
I312 Detector start Start-up of the detector
I313 Date/Time update Date or time modification
I314 Cell firmware update Analyzer cell firmware update
I315 CPU firmware update Detector firmware update
I316 LCD firmware update Control panel firmware update
I317 Voice update Voice update
I318 Full parameters reset Complete detector parameter reset
I319 Filament change Filament change from the Maintenance menu
I321 Storage delay Detector switched off for 15 days (minimum)
Calibrations history
The calibrations history records the calibrations made.
Access: Menu [Maintenance] [History] [Calibrations history]