A Roughing C Normal mode
B Gross Leak mode D High sensitivity mode
1 Set point for switching to Gross Leak mode 3 Set point for switching to High sensitivity mode
2 Set point for switching to Normal mode 4 Pressure
Access: Menu [Test] [Mode]
Choice - Setting limit
1)
Mode To be selected Gross Leak
Normal
High sensitivity
1) Initial setting: see chapter “Tree diagram to the Settings menu”
For quick access from the main screen, configure a function key to [MODE] (see chapter
“Function keys”).
By default, the leak detector is set to work with 'hard vacuum' test method and 'normal' test
mode: this setting meets the majority of users’ needs.
8.2.3 Type of probe
This menu is used to select the sniffer probe mode used in sniffer (see chapter “Accessories”).
Access: Menu [Test] [Probe type] Choice - Setting limit
Probe type Read only
Standard sniffer probe: model with rigid nozzle only
Standard
Set the probe clogged set point to verify that the sniffer probe is operational (see chapter
“Set points”).
8.2.4 Cycle end
This function allows automatic check of the roughing time and test timer in a hard vacuum test.
Settings menu
54/98