11 Monitoring NBP Calibrating NBP
160
Calibrating NBP
NBP is not user-calibrated. Cuff-pressure transducers must be verified and calibrated, if necessary, at
least once every two years by a qualified service professional. See the Service Guide for details.
(60)
Cuff pressure
Venous puncture measurement mode
VP
NBP
01:45
Time left in venous