EasyManuals Logo

Philips IntelliVue X2 User Manual

Philips IntelliVue X2
308 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #194 background imageLoading...
Page #194 background image
15 Assigning Two Devices to One Patient Functions Available For Devices Connected Via SRR
186
Refer to the Information Center Instructions for Use for more details on the Suspend/Pause alarms
behavior of the Information Center and telemetry device.
Using Standby
When you select Standby mode at the monitor, the bedside goes into Standby mode but the telemetry
device will continue monitoring.
Refer to the Information Center Instructions for Use for details on the Standby behavior of the
Information Center and telemetry device.
Functions Available For Devices Connected Via
SRR
The Telemetry Device window can be opened from the Measurement Selection
window, by selecting the Tele pop-up key, or via Main Setup -> Telemetry Device.
The window shows the equipment label of the assigned telemetry device and, for monitors with SRR
capability, a battery status symbol and a symbol indicating the signal quality of the
SRR link .
If the monitor has a connection to an Information Center, the pop-up keys Unpair to Mon. and
Unpair to Tele let you end the device pairing. After unpairing, the Information Center will
receive data exclusively from the monitor or from the telemetry device.
If the monitor does not have a connection to an Information Center, or there is a pairing mismatch,
the Unassign Tele pop-up key lets you end the telemetry device assignment and close the SRR
link.
Temporarily Stopping the Short Range Radio Connection
To temporarily disable the SRR connection for this device, for example, in case of interference:
♦ Select the Stop SRR pop-up key.
This closes the SRR link and the telemetry device switches over to standard telemetry transmission.
The battery status symbol will no longer be shown and the SRR quality indicator symbol will be
crossed out.
To return to SRR use:
♦ Select the Start SRR pop-up key and press the Check button on the telemetry transceiver.
Telemetry Device
Tele 1

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Philips IntelliVue X2 and is the answer not in the manual?

Philips IntelliVue X2 Specifications

General IconGeneral
BrandPhilips
ModelIntelliVue X2
CategoryControl Unit
LanguageEnglish

Related product manuals