25 Default Settings Appendix Cardiac Output Default Settings
290
Cardiac Output Default Settings
CO
2
Default Settings
C.O. Settings
Factory Defaults
Adult Pedi Neo
Auto-Calibration On
Tblood High Limit 39.0 ºC
Tblood Low Limit 36.0 ºC
Tblood Alarm On/Off On
Te m pe r a tu r e Un i t î™±C
Color Green
CCO/CCI Settings Factory Adult Factory Pedi Factory Neo
Settings common to CCO and CCI
Alarms from CCO
CCO from ABP
Color Green
CCO
Alarms On
CCO High Limit 8.5 l/min 3.7 l/min 1.3 l/min
CCO Low Limit 4.0 l/min 2.6 l/min 0.3 l/min
CCI
Alarms On
CCI High Limit 4.3 l/min/m
2
3.7 l/min/m
2
5.2 l/min/m
2
CCI Low Limit 2.0 l/min/m
2
2.6 l/min/m
2
1.2 l/min/m
2
CO
2
Settings
Factory Defaults
Adult Pedi Neo
etCO
2
low 30
etCO
2
high 50
imCO
2
high 4
CO
2
Alarms on
Unit mmHg
Scale 40 mmHg
ImCO
2
on
N
2
O Corr (only M3015A/M3016A) Off
Oxygen Corr (only M3014A) 16%
Gas Corr (only M3014A) Off
Humidity Corr BTPS