Invasive Pressure Default Settings 25 Default Settings Appendix
289
PAP Settings
Factory Defaults
Adult Pedi Neo
Alarms from Dia.
High Limit 34/16 (20) 60/4 (26) 60/4 (26)
Low Limit 10/0 (0) 24/-4 (12) 24/-4 (12)
Alarms On
Extreme Alarms Enabled
Enabled Enabled
Extreme High 5 5 5
Extreme Low 5 5 5
High Clamp 45/20 (25) 65/5 (35) 65/5 (35)
Low Clamp 5/-5 (-5) 15/-5 (5) 15/-5 (5)
Scale 30
Speed 25 mm/s
Mean Only No
Filter 12 Hz
Mercury Cal. Yes
Artifact Suppr. 60 sec
Unit mmHg
Color Yellow
ICP, IC1, IC2 Settings
Factory Defaults
Adult Pedi Neo
Alarms from Mean
High Limit 14/6 (10) 10/2 (4) 10/2 (4)
Low Limit 6/-4 (0) 2/-4 (0) 2/-4 (0)
Alarms On
Extreme Alarms Enabled
Enabled Enabled
Extreme High 10 10 10
Extreme Low 10 10 10
Low Clamp 20/10 (-5) 15/5 (-5) 15/5 (-5)
High Clamp 0/-5 (0) 0/-5 (0) 0/-5 (0)
Scale 30
Speed 25 mm/s
Mean Only Yes
Filter 12 Hz
Mercury Cal. Yes
Artifact Suppr. 60 sec
Unit mmHg
Color Magenta