25 Default Settings Appendix Invasive Pressure Default Settings
288
Extreme High 15
10 5
Extreme Low 15
10 5
High Clamp 190/100 (125) 140/80 (100) 105/75 (75)
Low Clamp 80/45 (65) 60/35 (45) 45/15 (30)
Scale 150 100 100
Speed 25 mm/s
Mean Only No
Filter 12 Hz
Mercury Cal. Yes
Artifact Suppr. 60 sec
Unit mmHg
Color Red
CVP, RAP, LAP, UVP Settings
Factory Defaults
Adult Pedi Neo
Alarms from Mean
High Limit 14/6 (10) 10/2 (4) 10/2 (4)
Low Limit 6/-4 (0) 2/-4 (0) 2/-4 (0)
Alarms On
Extreme Alarms Enabled
Extreme High 5 5 5
Extreme Low 5 5 5
High Clamp 20/10 (15) 15/5 (10) 15/5 (10)
Low Clamp 0/-5 (-5) 0/-5 (-5) 0/-5 (-5)
Scale 30
Speed 25 mm/s
Mean Only Yes
Filter 12 Hz
Mercury Cal. Yes
Artifact Suppr. 60 sec
Unit mmHg
Color light blue (cyan)
Average in Mon. No
Signal Quality On
ABP, ART, Ao, BAP, FAP, P, P1, P2, P3,
P4, UAP Settings
Factory Defaults
Adult Pedi Neo