25 Default Settings Appendix ECG, Arrhythmia, ST and QT Default Settings
282
ECG, Arrhythmia, ST and QT Default Settings
ECG Settings
Factory Defaults
Adult Pedi Neo
High Limit 120 bpm 160 bpm 200 bpm
Low Limit 50 bpm 75 bpm 100 bpm
Alarms On
Alarm Source Auto
ECG On
QRS Volume 1
Primary Lead II
Secondary Lead V
Analysis Mode Multi-lead
Lead Placement Standard
Mod. Lead Placment Off
Filter Monitor
Speed 25 mm/s
Auto Filter Off
Default ECG Size x1
Color Green
Asystole Thresh 4.0 sec 3.0 sec
ExtrTachy 20 bpm
Tachy Clamp 200 bpm 220 bpm 240 bpm
ExtrBrady 20 bpm
Brady Clamp 40 bpm 50 bpm
ALL ECG IN. On
Fallback On
Alarms Off Enabled
Alarm Source Sel. Enabled
Va Lead V2
Vb Lead V5
SyncPulse sensit Medium
SyncPulse Marker On
Fallback ECGTele Enabled
Arrhythmia Settings
Factory Defaults
Adult Pedi Neo
Arrhythmia
On Off
Pause Threshold 2.0 sec 1.5 sec
VTach HR 100 120 150
VTach Run 5