NBP Default Settings 25 Default Settings Appendix
287
NBP Default Settings
Te m p e r a t u r e D e f a u l t S e t t i n g s
Invasive Pressure Default Settings
NBP Settings
Factory Defaults
Adult Pedi Neo
Mode Auto Manual
Alarms from Sys.
High Alarm Limit 160/90 (110) 120/70 (90) 90/60 (70)
Low Alarm Limit 90/50 (60) 70/40 (50) 40/20 (24)
Alarms On
NBP On
Repetition Time 10 min
Pulse (NBP) On
Unit mmHg
Done Tone Off
Start Time Synchronized
VP Pressure 60 mmHg 40 mmHg 30 mmHg
Reference Auscultatory Invasive
Color Red
Temp Settings
Factory Defaults
Adult Pedi Neo
Low Limit 36
High Limit 39
Alarms On
Unit î™±C
Range 35...43
Color Green
ABP, ART, Ao, BAP, FAP, P, P1, P2, P3,
P4, UAP Settings
Factory Defaults
Adult Pedi Neo
Alarms from Sys.
High Limit 160/90 (110) 120/70 (90) 90/60 (70)
Low Limit 90/50 (70) 70/40 (50) 55/20 (36)
Alarms On
Extreme Alarms Disabled