24 Specifications Monitor Performance Specifications
254
Monitor Performance Specifications
Altitude Range Operating -500 m to 3000 m (10000 ft)
Storage including
transportation
-500 m to 4600 m
1
(15000 ft)
Ingress Protection Monitor IP32 (protected against ingress of water when the water is
dripping vertically and the monitor is tilted up to 15°)
External Power Supply
(M8023A)
IP31(protected against the ingress of solid foreign objects 2.5 mm
in diameter or larger, and the ingress of water when the water is
dripping vertically) when rested on its rubber feet on a flat, level
surface.
IP32 when mounted as shown in the section ““Mounting the
External Power Supply (M8023A)” on page 3 of the Installation
chapter and as described in the Service Guide.
1.Sufficient for flight altitudes up to 12,000 m with pressurized cabins.
Item Condition Range
X2 (M3002A) Performance Specifications
Power Specifications Power consumption <12 W average
<30 W while battery is charging
Operating Voltage
36 to 60 V DC floating
Battery Specifications Operating Time
(with new, fully
charged battery at
25°C)
Basic monitoring configuration: 2.5 hours
(Brightness set to Optimum, ECG/Resp, SpO
2
measurements in use, NBP measurement every 15 minutes)
Charge Time
When X2 is off: 2 hours
When X2 is in use, and connected to MP20/30/40/50/60/70/
80/90 without extensions: 12 hours approx.
When X2 is in use, and connected to the external power
supply (M8023A), without extensions: 12 hours approx.
Indicators Alarms Off red (crossed-out alarm symbol) LED
Alarms
red/yellow/light blue (cyan) LED
On/Standby / Error
green / red LED
AC Power
green LED
Battery
yellow (charging)/red blinking (empty) LED
External Power
green LED
Sounds Audible feedback for user input
Prompt tone
QRS tone, or SpO
2
modulation tone
3 different alarm sounds
Tr e n d s Resolution 12 or 16 numerics @ 12 sec, 1 minute, 5 minute resolution.
Information
Multiple choices of number of numerics, resolution and
duration depending on trend option and application area.
For example:
For neonatal, you can choose between 12 and 16 numerics.