19 Printing Patient Reports Stopping Reports Printouts
208
Stopping Reports Printouts
To stop Reports printing, in the Reports menu, select
Delete Report to stop the current print job
Delete All Repts to cancel all queued report printouts
Scheduled Rep. to toggle to Off and switch off scheduled reports.
Setting Up Reports
ECG report layout and Auto Reports settings must be defined in Configuration Mode.
The content you define in the individual Setup Reports menus will be used in all reports of this
type: for example, if you set a print speed of 50 mm/sec in the ECG Reports menu, every ECG
report will be printed at this speed, irrespective of format.
Setting Up ECG Reports
The settings you choose in the ECG Reports menu apply for all ECG reports printed. To set up
ECG Reports,
1 Select Main Setup -> Reports -> ECG Reports. Note the report layout configured for
your monitor, either Internat. or Cabrera. This
setting is inactive (“grayed-out”) in Monitoring Mode
and can only be changed in Configuration Mode.
2
Select ECG Gain and set the required ECG Gain to define how ECG waves will appear on the
ECG report printouts.
3 Select Speed and select the report print speed.
Report types Report contents How to start printing reports
Vital Signs Report depends on selected trend group, data
resolution, and period.
In the Vital Signs window, select Print
Graphic Trends Report In the Graphic Trends window, select Print
ECG Reports Depends on format selected Select the ECG Report SmartKeys, if configured
ST Map Report ST data in a transversal and/or horizontal
multiaxis diagram,
In the ST Map window, select Print.
Alarm Limits Report Graphic and numeric report of all
current alarm limits
In the Alarm Limits window, select Print
Limits
Realtime Reports Patient data and numerics, and either:
all displayed waves OR all measured
waves OR all measured RT waves
Select the Realtime Reports SmartKey, if
configured
Battery Report Battery serial number, information on
the currently-measured and potential
charge status of the battery
In the Battery Status window, select
Print Status.
Central Report A/B/C These are reports configured at the
Information Center. Monitors connected
via the IntelliVue Instrument Telemetry
wireless network can have these reports
available.
In the Main Setup select Reports then
Central Report A, B or C.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Philips IntelliVue X2 and is the answer not in the manual?

Philips IntelliVue X2 Specifications

General IconGeneral
BrandPhilips
ModelIntelliVue X2
CategoryControl Unit
LanguageEnglish

Related product manuals