Measurement Specifications 24 Specifications
263
NBP
Complies with IEC 60601-2-30:1999/EN60601-2-30:2000.
SpO
2
Alarm
Specifications Range Adjustment Delay
SpO
2
Adult: 50 to 100%
Pedi/Neo: 30 to 100%
1% steps (0, 1, 2, 3,... 30)
+4 seconds
Desat Adult: 50 to Low alarm limit
Pedi/Neo: 30 to Low alarm limit
1% steps
Pulse 30 to 300 bpm Adult:
1bpm steps (30 to 40bpm)
5 bpm steps (40 to 300 bpm)
Pedi/Neo:
1bpm steps (30 to 50bpm)
5 bpm steps (50 to 300 bpm)
max. 14 seconds
Tac hyc ardia Difference to high limit 0 to 50 bpm 5bpm steps max. 14seconds
Clamping at 150 to 300 bpm
5bpm steps
Bradycardia Difference to low limit 0 to 50 bpm 5bpm steps max. 14seconds
Clamping at 30 to 100 bpm
5bpm steps
NBP Performance Specifications
Measurement Ranges Systolic Adult: 30 to 270 mmHg (4 to 36 kPa)
Pedi: 30 to 180 mmHg (4 to 24 kPa)
Neo: 30 to 130 mmHg (4 to 17 kPa)
Diastolic
Adult: 10 to 245 mmHg (1.5 to 32 kPa)
Pedi: 10 to 150 mmHg (1.5 to 20 kPa)
Neo: 10 to 100 mmHg (1.5 to 13 kPa)
Mean
Adult: 20 to 255 mmHg (2.5 to 34 kPa)
Pedi: 20 to 160 mmHg (2.5 to 21 kPa)
Neo: 20 to 120 mmHg (2.5 to 16 kPa)
Pulse Rate
Adult: 40 to 300
Pedi: 40 to 300
Neo: 40 to 300
Accuracy Max. Std. Deviation: 8 mmHg (1.1 kPa)
Max. Mean Error: ±5 mmHg (±0.7 kPa)
Pulse Rate Measurement Accuracy 40 to 100 bpm: ±5 bpm
101 to 200 bpm:
±5% of reading
201 to 300 bpm:
±10% of reading
(average over NBP measurement cycle)
Pulse Rate Range 40 to 300 bpm
Measurement Time Typ ic al a t HR > 60 bp m
Auto/manual/sequence: 30 seconds (adult)
25 seconds (neonatal)
Stat: 20 seconds
Maximum time: 180 seconds (adult/pediatric)
90 seconds (neonates)