25 Default Settings Appendix ECG, Arrhythmia, ST and QT Default Settings
284
Lead-independent ST Settings
Factory Defaults
Adult Pedi Neo
ST Alarm Mode Single ST
Alarms On
ST Analysis On Off
ST-Index On
ISO Point -80 ms
J Point 48 ms
ST Point J+60
Lead I, II, III, V, aVR, aVL, aVF, V1-6,
MCL Settings
Factory Defaults
Factory Adult Factory Pedi Factory Neo
ST
(Label) On Off
For Alarm Mode = Single-ST
ST(Label) High +2.0 mm
ST
(Label) Low -2.0 mm
For Alarm Mode = Multi-ST
ST
(Label) High +1.0 mm
ST
(Label) Low -1.0 mm
QT Settings
Factory Defaults
Factory Adult Factory Pedi Factory Neo
QT Lead All
QTc High Limit 500 ms 480 ms 460 ms
QTc Limit 60 ms
QTc High Alarm On/Off On
QTc High Alarm On/Off On
QT Analysis Off
QTc Formula Bazett