Te c h n i c a l A l a r m M e s s a g e s ( I N O P s ) 5Patient Alarms and INOPs
65
CCO NO <Pressure label>
numeric is replaced by -?-
INOP tone may sound
C.O. CCO/CCI cannot be calculated. Make sure that the pressure
chosen in the
Setup CCO menu under CCO From
matches the pressure measured with the arterial catheter for
CCO measurement. A pressure from an external device cannot
be used. Select another pressure label, either ABP, Ao, ART,
BAP, FAP, or UAP.
CCO NO CALIBRATION
numeric is replaced by -?-
C.O. The CCO measurement is currently not calibrated.
CCO NO PRESS
at Information Center
C.O. CCO/CCI cannot be calculated. Make sure that the pressure
chosen in the
Setup CCO menu under CCO From
matches the pressure measured with the arterial catheter for
CCO measurement. A pressure from an external device cannot
be used. Select another pressure label, either ABP, Ao, ART,
BAP, FAP, or UAP.
CCO NOT SUPPORTED
numeric is replaced by -?-
INOP tone
C.O. A catheter for transpulmonary C.O. measurements has been
unplugged and replaced with a Right Heart C.O. catheter, or
the measurement mode has been changed manually. Silencing
this INOP switches the measurement off.
CCO/CCI OVERRANGE
numeric is replaced by -?-
INOP tone
C.O. The measured CCO or CCI value is not within the specified
range for CCO/CCI measurement.
CCO <Pressure label>
INVALID
numeric is replaced by -?-
INOP tone may sound
C.O. The arterial pressure selected for pulse contour calculation for
CCO is available but currently invalid. Make sure the pressure
transducer is connected and the zero calibration is valid.
CCO PRESS INVALID
at Information Center
C.O. The arterial pressure selected for pulse contour calculation for
CCO is available but currently invalid. Make sure the pressure
transducer is connected and the zero calibration is valid.
CCO PRESS OVERRANG
numeric is replaced by -?-
INOP tone
C.O. The mean value of the arterial pressure values used for pulse
contour calculation for CCO is below 0 mmHg or above 300
mmHg.
CCO PULSE OVERRANG
numeric is replaced by -?-
INOP tone
C.O. The pulse rate of the pressure used for pulse contour
calculation for CCO is below 30 bpm or above 240 bpm.
CCO/Tbl NO TRANSD
Numeric is replaced by -?-
INOP tone
C.O. No transducer attached to the module or catheter
disconnected.
CCO RECALIBRATE
numeric is replaced by -?-
C.O. The most recent CCO or CCI calibration was made over 8
hours ago or the arterial pressure measurement used for CCO
calculation has been zeroed after the CCO calibration was
performed. You should recalibrate CCO or CCI with
transpulmonary C.O. measurements at least every 8 hours or
when the hemodynamic condition of the patient has changed.
The pressure measurement must be zeroed before a CCO
calibration.
CENTRAL:TELE ONLY
INOP tone
Monitor System connectivity via telemetry device is limited (No alarms,
only local numerics) when in companion mode and host
monitor does not have system connectivity. Only telemetry
device parameters can be displayed at central station.
Charge BATT1/BATT2 now
INOP tone
Batteries Battery must be charged. Connect the monitor to mains power
or exchange the battery.
INOP Message, Indication Source What to do
To Purchase, Visit Avobus.com or call 1-800-674-3655

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Philips IntelliVue X2 and is the answer not in the manual?

Philips IntelliVue X2 Specifications

General IconGeneral
BrandPhilips
ModelIntelliVue X2
CategoryControl Unit
LanguageEnglish

Related product manuals