Te c h n i c a l A l a r m M e s s a g e s ( I N O P s ) 5Patient Alarms and INOPs
75
<pTemp label> CHECK PROBE
Numeric is replaced by -?-
INOP tone.
Predictive
Te m p
Check that the probe holder is correctly installed and that a
compatible probe is in use. If the INOP does not clear, remove
the probe from the holder then replace it. If the INOP still does
not clear, disconnect and reconnect the probe.
<pTemp label> DEACTIVATED
INOP tone.
Predictive
Te m p
The Predictive temperature has been deactivated.
<pTemp label> EQUIP MALF
Numeric is replaced by -?-
INOP tone.
Predictive
Te m p
The Predictive Temperature hardware is defective. Contact
your service personnel.
<pTemp label> INCOMPAT. Predictive
Te m p
The Predictive Temperature version is not supported by the
monitor.
<pTemp label> MEAS FAILED
Numeric is replaced by -?-
INOP tone.
Predictive
Te m p
The ambient temperature is outside the specified range (10 to
40°C /50 to 104°F). Bring the temperature into range to
continue monitoring. If the INOP does not clear, disconnect
and reconnect the probe.
<pTemp label> NO PROBE
Numeric is replaced by -?-
INOP tone.
Predictive
Te m p
Connect a probe to the Predictive temperature unit.
<pTemp label> OVERRANGE
Numeric is replaced by -?-
INOP tone.
Predictive
Te m p
The temperature at the measurement site is out of range. Check
that the probe is correctly located. If the INOP does not clear,
disconnect and reconnect the probe.
<pTemp label> PROBE MALF
Numeric is replaced by -?-
INOP tone.
Predictive
Te m p
The connected probe may be defective. Disconnect and
reconnect the probe. If the INOP does not clear, try another
probe.
pTaxil INOPS Predictive
Te m p
See <pTemp label> INOPS
pToral INOPS Predictive
Te m p
See <pTemp label> INOPS
pTrect INOPS Predictive
Te m p
See <pTemp label> INOPS
(!!)PW:Action Required Protocol
Watch
The protocol currently running requires a user response. Check
which pop-up window is displayed and provide the appropriate
response.
PW:Check Settings Protocol
Watch
Contact your service personnel. Settings could not be loaded or
interpreted correctly
PW in conflict Protocol
Watch
There is a patient information mismatch which has not yet
been resolved (>15 minutes).
RA LEAD OFF
Numeric is replaced by -?-
INOP tone.
ECG The RA electrode has become detached from the patient or the
lead set has been changed. Reattach the electrode or select
New
Lead Setup
in the Setup ECG menu to confirm the
new lead set.
RAP INOPS PRESS See <Pressure label> INOPS (under Pressure).
Rem.Alarmdev.Malf
INOP tone
Monitor There is a problem with the connection to the remote alert
device. Contact your service personnel to check the remote
alert device and its connections.
Rem. Display Malf. Remote
display
There is a problem with an input device at the remote display.
Perform a visual and functional check of all input devices.
Contact your service personnel.