5Patient Alarms and INOPs Te c h n i c a l A l a r m M e s s a g e s ( I N O P s )
78
SPIRO NO BREATH Spirometry No breath was detected for more than 25 seconds. Breath
derived numerics are not available.
SPIRO GAS COMPENS? Spirometry Gas compensation is set togas analyzer” but not all gases
necessary for compensation are measured by a gas monitor.
Some of the fall-back values provided by the user are used.
Measurement accuracy might be reduced.
SPIRO PATIENT CAT. Spirometry Mismatch of patient size configured in the host monitor and
sensor type plugged into the module. Check the instructions
on selecting the correct sensor in the Spirometry chapter.
SpO
2
CHK SOURCES
Numeric is replaced by -?-
SpO
2
Difference
Not all measurements or values required to perform the
calculation are available. Check measurement sources.
SpO
2
CHK UNITS
Numeric is replaced by -?-
SpO
2
Difference
The monitor has detected a conflict in the units used for this
calculation. Check the unit settings.
<SpO
2
label> DEACTIVATED
INOP tone
SpO
2
The SpO
2
measurement label in the measurement device has
been deactivated by deactivating the label in the Measurement
Selection window. The measurement automatically disappears
from the display. To switch the measurement on again,
reactivate the measurement label in the Measurement Selection
window.
<SpO
2
label> EQUIP MALF
Numeric is replaced by -?-
INOP tone.
SpO
2
The MMS or module is faulty. Unplug and replug the MMS or
module. If the INOP persists, contact your service personnel.
<SpO
2
label> ERRATIC
Numeric is replaced by -?-
INOP tone.
SpO
2
Check the sensor placement. Try another adapter cable and
sensor. If the INOP persists, contact your service personnel.
<SpO
2
label> EXTD. UPDATE
Label is displayed with a -?-
(questionable numeric)
SpO
2
The update period of displayed values is extended due to an
NBP measurement on the same limb or an excessively noisy
signal.
<SpO
2
label> INTERFERNCE
Numeric is replaced by -?-
INOP tone.
SpO
2
There is too much interference, caused by a high level of
ambient light and/or electrical interference. Cover the sensor to
minimize ambient light. If the INOP persists, make sure that
the sensor cable is not damaged or positioned too close to
power cables.
<SpO
2
label> LOW PERF
Label is displayed with a -?-
(questionable numeric)
SpO
2
Accuracy may be compromised due to very low perfusion.
Stimulate circulation at sensor site. If INOP persists, change
the measurement site.
<SpO
2
label> NOISY SIGN.
Numeric is replaced by -?-
INOP tone.
SpO
2
Excessive patient movement or electrical interference is causing
irregular pulse patterns. Try to reduce patient movement or to
relieve the cable strain on the sensor.
<SpO
2
label> NON-PULSAT.
Numeric is replaced by -?-
INOP tone.
SpO
2
Check the perfusion at measurement site. If necessary,
stimulate circulation or change measurement site. If the INOP
is due to NBP measurement on the same limb, wait until the
NBP measurement is finished.
<SpO
2
label> NO SENSOR
Numeric is replaced by -?-
INOP tone.
SpO
2
Make sure the SpO
2
sensor is connected. If the INOP persists,
try another adapter cable and sensor. If you silence this INOP,
the measurement will be switched off.
<SpO
2
LABEL> POOR SIGNAL
Label is displayed with a -?-
(questionable numeric)
SpO
2
The signal condition of the SpO
2
measurement is poor and
measurement accuracy may be compromised.
INOP Message, Indication Source What to do

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Philips IntelliVue X2 and is the answer not in the manual?

Philips IntelliVue X2 Specifications

General IconGeneral
BrandPhilips
ModelIntelliVue X2
CategoryControl Unit
LanguageEnglish

Related product manuals