5Patient Alarms and INOPs Te c h n i c a l A l a r m M e s s a g e s ( I N O P s )
80
Tblood NO TRANSDUC
Numeric is replaced by -?-
INOP tone
C.O. No transducer attached to the module or catheter
disconnected.
Tblood OVERRANGE
Numeric is replaced by -?-
C.O. Tblood out of range 17°C - 43°C.
Tcereb INOPS TEMP See <Temp label> INOPs (under Temp)
Tcore INOPs TEMP See <Temp label> INOPs (under Temp).
tcpO
2
(or tcpCO
2
or tcGas)
CAL FAILED
Numeric is replaced by -?-
INOP tone.
tcGas A calibration failed. Check the calibration unit, gas pressure,
and tubing connections, then restart the calibration. If the
calibration has failed more than once, remembrane the
transducer and restart the calibration. If this INOP persists,
contact your service personnel.
tcpO
2
(or tcpCO
2
or tcGas)
CAL REQUIRD
Numeric is replaced by -?-
INOP tone.
tcGas Calibration is required before applying the transducer to the
patient.
Insert a membraned transducer into the calibration chamber on
the module, connect the calibration unit to the calibration
chamber, open the gas valve and start the calibration. If this
INOP occurs during a calibration, there may be a module or
transducer malfunction: contact your service personnel.
tcpO
2
(or tcpCO
2
or tcGas)
CAL RUNNING
Numeric displays first -?- , then
numeric is displayed with a ?
tcGas Wait until the tcpO
2
/tcpCO
2
calibration is finished.
tcpO
2
(or tcpCO
2
or tcGas)
CHECK TIME
tcGas Site Timer due to time out in 15 minutes or less.
tcpO
2
(or tcpCO
2
or tcGas)
CHANGE SITE
If Heat Switch Off is configured to Yes,
numeric is replaced by -?-
INOP tone.
tcGas Site Timer has timed out. Change the application site to avoid
skin burns. To reset the Site Timer, either calibrate and change
the measurement site, or change the measurement site and reset
the Site Timer manually by selecting the appropriate site time
from the
Setup TCGas menu.
tcpO
2
(or tcpCO
2
or tcGas)
EQUIP MALF
Numeric is replaced by -?-
INOP tone.
tcGas There is a malfunction in the transducer or module. Connect
another transducer. If this INOP persists, contact your service
personnel.
tcpO
2
(or tcpCO
2
or tcGas)
NO TRANSDUC
Numeric is replaced by -?-
INOP tone.
tcGas No transducer is connected to the tcpO
2
/tcpCO
2
module.
Silencing the alarm switches off the measurement.
tcpO
2
(or tcpCO
2
or tcGas)
STABILIZING
Numeric is displayed with a ?
tcGas The transducer has not yet reached the selected temperature
and/or skin hyperemization is not yet finished. This INOP will
disappear within three minutes.
tcpO
2
(or tcpCO
2
or tcGas)
UNPLUGGED
Numeric is replaced by -?-
INOP tone.
tcGas The measurement is switched on but the module is unplugged.
The measurement automatically disappears from the display.
Silencing this INOP switches off the measurement.
TELE CONFIG UNSUPP
INOP tone
Monitor Telemetry device not supported (companion mode)
INOP Message, Indication Source What to do