Tr a n s ferr ing Pati ents 6Managing Patients
89
Transferring a Centrally-Monitored Patient
WARNING Measurements from a MMS extension connected to an X2 are not available when the X2 is running on
battery power. They are only available when the X2 is powered from AC mains, either when connected
to a host monitor or the external power supply (M8023A).
Scenario: A centrally-monitored patient is moved to another monitoring location on the same
Information Center database server without interrupting the collection of patient trend information,
using the X2 as both MMS and transport monitor.
1 Before you disconnect the X2 from the host monitor, select the patient name field or select the
Admit/Dischrge SmartKey on the host monitor to open the Patient Demographics
window, then select the Transfer pop-up key. If the patient is not admitted or not monitored
by an Information Center, the Transfer key is inactive (“grayed-out”).
This step preserves the patient’s demographic data during the transfer.
2 Remove the X2 and any connected extensions from the host monitor.
3 Move the patient using the X2 as the transport monitor.
4 At the new location, connect the X2 to the new host monitor. If the host monitor detects a patient
mismatch, a window will open showing your patient’s data and asking Complete transfer
of this patient?.
5 Select Yes to complete the transfer.This re-admits the patient from the transfer list to the new
monitor. This will upload the patient demographics, and, if configured, the measurement settings
and trend data stored in the X2 to the receiving host monitor.
6 Veri fy that the setting s for patient categ or y and paced mode are correc t.
Transferring a Centrally-Monitored Patient using IIT
WARNING Measurements from a MMS extension connected to an X2 are not available when the X2 is running on
battery power. They are only available when the X2 is running on external power, either when
connected to a host monitor or the external power supply (M8023A).
Scenario: A centrally-monitored patient is moved with an X2 to another monitoring location on the
same Information Center database server without interrupting the collection of patient trend
information. The X2 has an IntelliVue Instrument Telemetry interface (IIT), is connected to a host
monitor, and is declared as a telemetry device at the Information Center.
NOTE The Transfer key is not available while the X2 is connected to a host monitor (Companion
Mode is indicated).
1 Remove the X2 and any connected extensions from the host monitor.
2 Move the patient using the X2 as the transport monitor.
3 At the new location, just before connecting the X2 to the new host monitor:
–open the Patient Demographics window.
–select the Transfer pop-up key.
wait until the transfer has completed.