AFV
+
Series High Power Programmable AC Power Supply
6-115
6.8.6 LIMit:CURR
Function Description:Set the upper limit current value
Command Description:LIM:CURR 100.0\r\n
Parameter1:upper limit current value
Answer Command:N/A
Note
:
The set value of upper limit current cannot be greater than the rated current
value
6.8.7 LIMit:POWer
Function Description:Set the upper limit power value
Command Description:LIM:POWer 100.0\r\n
Parameter1:Upper limit power value
Answer Command:N/A
Note
:
The set value of upper limit power cannot be greater than the rated power
value
6.9 COMM Command
6.9.1 COMM:ERR?
Function Description :Read communication status of last command.
Command Description:COMM:ERR?\r\n
Parameter:N/A
Answer Command:COMM:ERR n\r\n
Parameter1:0—NO ERR, 1—NO End of String(\n), 2—Invalid Command