Inconsistent measurements:
1) Brackets incorrectly mounted: are the brackets slipping or loose
or mounted at an angle? (page 4-11). Were the brackets
accidentally knocked during or between measurements?
2) Measurement range too small or too few points: rotate through a
larger angle with more measurement points (page 4-33)
3) Machine preparation problem: excessive bearing play? Check
with Soft foot function (page 4-29). Is the foundation stable? If
suspect, split the coupling and measure with Pass mode (page
4-36). Is the coupling too stiff, e.g. membrane coupling? See
page 6-7.
4) Are the receiver and laser lenses clean? (page 6-36)
Difficulties in alignment:
1) Poor measurement, see reasons listed above.
2) Are corrections being made in the wrong direction? See the rules
on pages 4-34 and 4-50.
3) Check Soft foot (page 4-29)
4) Check machine preparation details (page 4-10). Is the foundation
stable? Are your horizontal moves reasonably accurate with jack
screws or similar?
Troubleshooting guide

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the PRÜFTECHNIK ROTALIGN ALI 3.500 and is the answer not in the manual?

PRÜFTECHNIK ROTALIGN ALI 3.500 Specifications

General IconGeneral
BrandPRÜFTECHNIK
ModelROTALIGN ALI 3.500
CategoryLaser Shaft Alignment Tools
LanguageEnglish

Related product manuals