MP3-PL Therapy Beam Analyzer T41050
Safety Information
12
D850.131.05/01 en
CAUTION
Improper handling.
Risk of injury!
If the water phantom is lowered to the bottom
position by means of the SCANLIFT lift mecha-
nism, only very little space (approx. 10 12 mm)
is left between the clamping screw of the cham-
ber cable holding device and the lower push bar
of the SCANLIFT.
While the lift element is moving, ensure that no
other persons are working on the SCANLIFT.
Otherwise, these persons are at risk of suffering
crush injuries at their hands and fingers.
CAUTION
Improper handling.
Risk of injury!
The chamber cable holding device is very heavy
(several kilograms).
Push the chamber cable holding device carefully
to its final position. Otherwise, there is a risk of
suffering crush injuries at the hands and fingers
at the end of the rails.
CAUTION
Improper handling.
Risk of injury!
When used with the SCANLIFT, rotating the
MP3-PL by means of the positioning device is
not allowed.
When the positioning device rotates, the base
plate of the chamber cable holding device may
collide with the higher push bar of the
SCANLIFT. This may damage the product.
While the lift element is moving, ensure that no
other person rotates the positioning device. Oth-
erwise, these persons are at risk of suffering
crush injuries at their hands and fingers.