ppbRAE 3000 User’s Guide
26
Must Calibrate. The user is prompted to calibrate the instrument
when calibration is due (as set by the calibration interval). There are
two programmable options:
Can’t Bypass. Unless calibration is performed, the
instrument cannot be used, and the only option is to turn off
the instrument.
Can Bypass. If calibration is due but the user does not want
to perform a calibration, the instrument can still be used. In
this case, the instrument records that the user has bypassed
the calibration requirement in a Policy Violation report.
Must Bump. The user is prompted to bump test the instrument when
a bump test is due (as set by the bump test interval). There are two
programmable options:
Can’t Bypass. Unless a bump test is performed, the
instrument cannot be used, and the only option is to turn off
the instrument.
Can Bypass. If a bump test is due but the user does not want
to perform one, the instrument can still be used. In this case,
the instrument records that the user has bypassed the bump
testing requirement in a Policy Violation report.
These are the screens that are shown on a ppbRAE 3000 after startup
if “Can Bypass” is selected: