-85-
S
S
C
C
Adjustment of installation altitude above sea level
SCC Gas from 04 - 2004
Note: The altitude settings of 0-499 and 500-999m are identical.
Therefore resetting of installation altitude needs to be done only when installing
above 1000m (3280ft) or below sea level (i.e. Red Sea)
A) - Switch unit ON
- Open control panel
- Set DIP switch 1 on PCB to „ON“ position
B) At Basic Settings select position 4 - Installation altitude
C) Press Dial key and select desired altitude.
D) Confi rm new altitude setting with „Store“ key
E) Switch unit OFF and ON again to store new setting!
F) To exit service program set DIP switch 1 to „OFF“ position
Perform a complete fl ue gas analysis at Max speed of each burner
at the „Function Test“ ,as well as a cross-check of CO2 at Min speed
of each burner at the „Function Test“