Secon 5
Integra 3™ Basic Operaon
22
Research Instruments Ltd
5
2. Take an empty and clean Petri dish (set-up dish), and lightly scratch a
cross in its centre, on the upper surface. This must be the same type
of Petri dish as used clinically.
3. Select a low magnicaon objecve (4x without spacer). Place the
Petri dish on the stage and use the microscope focus control to focus
on the cross. Do not adjust the microscope focus again and do not
remove the dish unl the pipee set-up is complete.
4. Select the set-up objecve (4x with spacer).
5. Li the Set-up Levers to the raised posion. Insert holding and injecon
pipees into the MPHs and clip the MPHs into the toolholders (PL3).
6. Adjust the angle of each pipee by turning the Pipee Angle Control.
7. Align the bend in each pipee by turning the Pipee Rotator unl the
toe of the pipee is in a straight line.
8. Use the Axial Drive to move the pipees so that the ps are close to
each other. Look into the eyepieces and bring the p of each pipee
to the centre of the eld of view. Use the Axial Drive, if necessary, to
adjust the pipee p sideways. Use the Centring Control to move the
pipee p forwards or backwards. The p might be out of focus - this
does not maer and will be adjusted next.