Secon 10
Integra 3™ Troubleshoong
55
Research Instruments Ltd
10



 






E09
HEATED STAGE
TEMPERATURE
OUTSIDE ±0.5°C
Medium
Heated Stage
Plate temperature
has deviated by
more than 0.5°C
from the setpoint
temperature.
Alarm enabled
5 minutes aer
the temperature
reaches ±2.5°
from the setpoint
temperature.
The Heated
Stage Plate
temperature
controller
connues to
operate.
This may be
caused by
placing either
hot or cold
objects on the
stage. In this
case either
remove the
object or wait
a short me for
the setpoint
temperature to
reached.
Sudden air
movements or
temperature
change can also
cause minor
temperature
uctuaons.
In this case
wait a short
while for the
temperature
controller to
respond.
Switch o the
Integra 3 and
verify that the
Heated Stage
Plate cables
and connectors
are secure and
undamaged.
Possible heater
or controller
fault.