EasyManua.ls Logo

Research Instruments INTEGRA 3 - Page 62

Default Icon
76 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Secon 10
Integra 3™ Troubleshoong
58
Research Instruments Ltd
10



 






E15
HEATED INSERT
TEMPERATURE
OUTSIDE ±0.5°C
Medium
Heated Stage
Insert temperature
has deviated by
more than 0.5°C
from the setpoint
temperature.
Alarm enabled
5 minutes aer
the temperature
reaches ±2.5°
from the setpoint
temperature.
The Heated
Stage Insert
temperature
controller
connues to
operate.
This may be
caused by
placing either
hot or cold
objects on the
stage. In this
case either
remove the
object or wait
a short me for
the setpoint
temperature to
reached.
Sudden air
movements or
temperature
change can also
cause minor
temperature
uctuaons.
In this case
wait a short
while for the
temperature
controller to
respond.
Switch o the
Integra 3 and
verify that the
Heated Stage
Insert cables
and connector
are secure and
undamaged.
Possible heater
or controller
fault.

Table of Contents