Secon 10
Integra 3™ Troubleshoong

Research Instruments Ltd
10



 






E20
HEATED STAGE
LOW HEATING
RATE
Medium
Heated Stage
Plate was not
able to achieve
the setpoint
temperature.
Alarm acvated
15 minutes aer
power on if the
temperature is
not within ±2.5°C
of the setpoint
temperature.
Temperature
controller
connues to
operate.
Switch o the
Integra 3 and
verify that the
Heated Stage
Plate cables
and connector
are secure and
undamaged.
Possible heater
or controller
fault. Operang
the
Integra 3 below
its ambient
temperature
specicaon
may produce
this alarm.
E21
THERMOSAFE
LOW HEATING
RATE
Medium
Thermosafe Air
Heang System was
not able to achieve
the setpoint
temperature.
Alarm acvated
15 minutes aer
power on if the
temperature is
not within ±2.5°C
of the setpoint
temperature.
Temperature
controller
connues to
operate.
Switch o the
Integra 3 and
verify that the
Thermosafe
Air Heang
System cables
and connector
are secure and
undamaged.
Possible heater
or controller
fault.
Operang the
Integra 3 below
its ambient
temperature
specicaon
may produce
this alarm.



 





