11
Digital Voltmeter (DVM) and Frequency
Counter
This series oscilloscope provides a built-in digital voltmeter (DVM) and frequency
counter, which enable you to perform more accurate measurements, improving user
experience in counter and frequency measurement.
11.1 Digital Voltmeter (DVM)
The built-in DVM of this oscilloscope provides 4-digit voltage measurements on any
analog channel. DVM measurements are asynchronous from the oscilloscope's
acquisition system and are always acquiring. You can enable the DVM measurements
in the following ways.
Click or tap the function navigation icon
> DVM to enable DVM
measurements.
Click or tap the
DVM button on the toolbar to enable DVM measurements.
Press the front-panel
key and then select DVM in the displayed "Analyse"
menu to enable the DVM measurements.
After the DVM measurements are enabled, the "DVM" label appears in the "Result"
bar at the right section of the screen, as shown in the figure below.
Source
Voltage Value
Mode
DVM Bandwidth
The voltage value in the label shows the measurement extrema over the last 3
seconds.
Click or tap the "DVM" label and then a window is displayed. Click or tap
Setting to
enter the DVM setting menu as shown in
Figure 11.1
. You can click or tap Remove to
disable DVM measurements.
11.1.1 Measurement Settings
After the DVM is enabled, click or tap the "DVM" label in the "Result" bar at the right
section of the screen and a window is displayed. Click or tap
Setting in the window to
enter the DVM setting menu, as shown in the figure below.
Digital Voltmeter (DVM) and Frequency Counter
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO800 User Guide
149

Table of Contents

Other manuals for Rigol DHO800 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DHO800 Series and is the answer not in the manual?

Rigol DHO800 Series Specifications

General IconGeneral
Channels2 or 4 analog channels
Vertical Resolution12-bit
Vertical Sensitivity1 mV/div to 10 V/div
Timebase Range1 ns/div to 1000 s/div
Power Supply100-240 VAC, 50/60 Hz
Waveform Capture Rate1, 000, 000 wfm/s
InterfacesUSB Host, USB Device, LAN
Digital ChannelsOptional 16-channel MSO
Protocols DecodingI2C, SPI, UART, CAN, LIN
Input Impedance1 MΩ
Trigger TypesEdge, Pulse, Video, Slope, Pattern, Duration, Timeout, Runt, Window