2 Command System RIGOL
MSO2000A/DS2000A Programming Guide
2-117
:LA:POD<n>:THReshold
Syntax
:LA:POD<n>:THReshold <threshold>
:LA:POD<n>:THReshold?
Description
Set or query the threshold of the specified default channel group. The default unit is V.
Parameter
Name Type Range Default
<n> Discrete {1|2} --
<threshold> Real -20.0V to +20.0V 0.7V
Return Format
The query returns the current threshold in scientific notation.
:LA:GROup<n>:APPend
Syntax
:LA:GROup<n>:APPend <digital>[,<digital>[,<digital>]...]
Description
Add channels into the specified user-defined channel group.
Parameter
Name Type Range Default
<n> Discrete {1|2|3|4} --
<digital> Discrete
{D0|D1|D2|D3|D4|D5|D6|D7|D8|D9|
D10|D11|D12|D13|D14|D15}
--