2 Command System RIGOL
MSO2000A/DS2000A Programming Guide
2-135
:MEASure:VAMP
:MEASure:VAVG
:MEASure:VBASe
:MEASure:VMAX
:MEASure:VMIN
:MEASure:VPP
:MEASure:VRMS
:MEASure:PVRMs
:MEASure:VTOP
:MEASure:MARea
:MEASure:MPARea
:MEASure:SOURce
Syntax
:MEASure:SOURce <sour>
:MEASure:SOURce?
Description
Set or query the signal source of the current measurement parameter.
Parameter
Name Type Range Default
<sour> Discrete
{D0|D1|D2|D3|D4|D5|D6|D7|D8|D9|D10|D11|D12|D13|D14|
D15|CHANnel1|CHANnel2|MATH}
CHANnel1
Explanation
Only channels currently turned on can be selected.
This command sets the signal source of all the measurement parameters except delay and phase.
The delay and phase measurements require two signal sources which can be set by
the
:MEASure:SETup:DSA and :MEASure:SETup:DSB, :MEASure:SETup:PSA
and
:MEASure:SETup:PSB commands respectively.
Return Format
The query returns D0, D1, D2, D3, D4, D5, D6, D7, D8, D9, D10, D11, D12, D13, D14, D15, CHAN1,
CHAN2 or MATH.

Table of Contents

Other manuals for Rigol DS2302A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DS2302A and is the answer not in the manual?

Rigol DS2302A Specifications

General IconGeneral
BrandRigol
ModelDS2302A
CategoryTest Equipment
LanguageEnglish

Related product manuals