2 Command System RIGOL
MSO2000A/DS2000A Programming Guide
2-141
:MEASure:SETup:MAX
:MEASure:SETup:MID
:MEASure:SETup:MIN
Syntax
:MEASure:SETup:MAX <max>
:MEASure:SETup:MAX?
:MEASure:SETup:MID <mid>
:MEASure:SETup:MID?
:MEASure:SETup:MIN <min>
:MEASure:SETup:MIN?
Description
Set or query the upper limit, middle value or lower limit of the auto measurement of the analog channels.
The unit is %.
Parameter
Name Type Range Default
<max> Integer 7 to 95 90
<mid> Integer 6 to 94 50
<min> Integer 5 to 93 10
Explanation
The upper limit, middle value and lower limit of the threshold level are expressed in percentage of
the amplitude and are used to define the time, delay and phase parameters. Setting the threshold
will affect all the time, delay and phase parameters. You can also use the
:MEASure:SETup:TYPE
command to select the threshold measurement setting.
The middle value set must be lower than the upper limit currently set and greater than the lower limit
currently set.
Return Format
The query returns an integer.

Table of Contents

Other manuals for Rigol DS2302A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DS2302A and is the answer not in the manual?

Rigol DS2302A Specifications

General IconGeneral
BrandRigol
ModelDS2302A
CategoryTest Equipment
LanguageEnglish

Related product manuals