RIGOL 2 Command System
2-172 MSO2000A/DS2000A Programming Guide
:MEASure:RTIMe
Syntax 1
:MEASure:RTIMe <chan>
Description 1
Enable the rise time measurement function of the specified channel.
Syntax 2
:MEASure:RTIMe? <chan>
Description 2
Enable the rise time measurement function of the specified channel and return the measurement result
(the unit is s).
Parameter
Name Type Range Default
<chan> Discrete {CHANnel1|CHANnel2|MATH} --
Return Format
The query returns the measurement result in scientific notation.
Example
:MEASure:RTIMe CHANnel1
:MEASure:RTIMe? CHANnel1
You can use the following commands to query the statistic maximum, minimum, current value, average
and standard deviation of the rise time measurement values of the specified channel waveform. Sending
the following commands will enable the statistic function automatically.
:MEASure:RTIMe:SMAXimum?
:MEASure:RTIMe:SMINimum?
:MEASure:RTIMe:SCURrent?
:MEASure:RTIMe:SAVerage?
:MEASure:RTIMe:SDEViation?