2 Command System RIGOL
MSO2000A/DS2000A Programming Guide
2-199
:MEASure:VTOP:SMAXimum?
:MEASure:VTOP:SMINimum?
:MEASure:VTOP:SCURrent?
:MEASure:VTOP:SAVerage?
:MEASure:VTOP:SDEViation?
Syntax 1
:MEASure:VTOP:SMAXimum <chan>
:MEASure:VTOP:SMINimum <chan>
:MEASure:VTOP:SCURrent <chan>
:MEASure:VTOP:SAVerage <chan>
:MEASure:VTOP:SDEViation <chan>
Description 1
Enable the amplitude top value measurement function of the specified channel and enable the statistic
function.
Syntax 2
:MEASure:VTOP:SMAXimum? <chan>
:MEASure:VTOP:SMINimum? <chan>
:MEASure:VTOP:SCURrent? <chan>
:MEASure:VTOP:SAVerage? <chan>
:MEASure:VTOP:SDEViation? <chan>
Description 2
Enable the amplitude top value measurement function of the specified channel; enable the statistic
function; return the maximum value, minimum value, current value, average value or standard deviation.
Parameter
Name Type Range Default
<chan> Discrete {CHANnel1|CHANnel2|MATH} --
Explanation
The unit is determined by the unit selected for the source channel (refer to the
:CHANnel<n>:UNITs
command).
Return Format
The query returns the measurement result in scientific notation.