Performing a Test
98
4 Using the Roche CARDIAC pipette, draw
exactly 150 µL (to the blue mark on the
pipette) heparinised blood from the blood
collection vial. Make sure the sample con-
tains no air bubbles.
For ordering information on Roche CARDIAC
pipettes see page 133.
5 Apply the entire sample to the application
area of the test strip.
Note: Always place the meter on a level,
vibration-free surface while applying the
sample until the necessary sample has been
absorbed completely by the test strip.
6 Touch to confirm that the sample has
been applied.
The hourglass icon appears as the meter
begins to process the sample. For better
control over the measurement process, you
should always confirm this yourself, instead of
waiting for the meter to detect the sample.
After you confirm that the sample has been
applied, or if the meter automatically recog-
nises that the sample has been applied, the
button disappears.
150 µL