Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
270User Manual 1173.9240.02 ─ 17
Parameters:
<State> ON | OFF
*RST: OFF
Example:
SENS:ADJ:SCAL:Y:AUTO ON
Manual operation: See "AF Auto Scale" on page 119
11.4.10 Configuring Standard Traces
Useful commands for trace configuration described elsewhere
DISPlay[:WINDow<n>]:TRACe:Y:SPACing on page 239
DISPlay[:WINDow<n>]:TRACe:Y[:SCALe] on page 238
Remote commands exclusive to trace configuration
DISPlay[:WINDow<n>]:TRACe<t>:MODE.........................................................................270
DISPlay[:WINDow<n>]:TRACe<t>:MODE:HCONtinuous....................................................271
DISPlay[:WINDow<n>]:TRACe<t>:SELect........................................................................272
DISPlay[:WINDow<n>]:TRACe<t>[:STATe].......................................................................272
[SENSe:]ADEMod:AM[:ABSolute][:TDOMain][:TYPE]........................................................ 272
[SENSe:]ADEMod:AM:RELative[:TDOMain][:TYPE].......................................................... 272
[SENSe:]ADEMod:AM:RELative:AFSPectrum[:TYPE]........................................................272
[SENSe:]ADEMod:FM[:TDOMain][:TYPE].........................................................................272
[SENSe:]ADEMod:FM:AFSPectrum[:TYPE]......................................................................272
[SENSe:]ADEMod:PM[:TDOMain][:TYPE]........................................................................ 272
[SENSe:]ADEMod:PM:AFSPectrum[:TYPE]......................................................................273
[SENSe:]ADEMod:SPECtrum[:TYPE]...............................................................................273
[SENSe:]AVERage:COUNt............................................................................................. 274
[SENSe:]AVERage<n>[:STATe<t>]..................................................................................275
[SENSe:]AVERage<n>:TYPE..........................................................................................275
[SENSe:][WINDow<n>:]DETector<trace>[:FUNCtion]........................................................ 275
[SENSe:][WINDow<n>:]DETector<t>[:FUNCtion]:AUTO.....................................................276
DISPlay[:WINDow<n>]:TRACe<t>:MODE <Mode>
This command selects the trace mode.
In case of max hold, min hold or average trace mode, you can set the number of single
measurements with [SENSe:]SWEep:COUNt. Note that synchronization to the end of
the measurement is possible only in single sweep mode.
In the Analog Demodulation application when you configure the traces for a window
with a specific evaluation (e.g. AM time domain), the traces in all windows with the
same evaluation are configured identically.
Configuring the Measurement