Instrument Functions
NRP2
110User Manual 1173.9140.02 ─ 07
"Normal"
Displays the frequencies of the primary and, if used, secondary chan-
nel.
The following information is additionally displayed to single-channel
power measurement results in time slots and time gates (with R&S
NRP-Z78x wideband sensors only):
"Pk, Pk/Av" main value Av
"Av, Pk/Av" main value Pk
"Av, Pk" main value Pk/Av
"Extremes"
Displays the maximum, the minimum and the max-min values since
the search for extreme values has been started.
"Statistics"
Displays the longterm mean, the standard deviation and the total
number of measurement results which have been evaluated since the
search for statistic values has been started.
Remote command:
CALCulate<[1]..8>:
DATA?
on page 225
CALCulate<[1]..8>:MAXimum:DATA? on page 232
CALCulate<[1]..8>:MINimum:DATA? on page 232
CALCulate<[1]..8>:PTPeak:DATA? on page 233
CALCulate<[1]..8>:MEAN:DATA? on page 232
CALCulate<[1]..8>:SDEViation:DATA? on page 234
CALCulate<[1]..8>:COUNt:DATA? on page 225
DISPlay[:WINDow]<[1]..4>:AVALue on page 241
4.4.4 Relative measurement
The "Relative" function allows you to relate a power or a power ratio to a reference
value. The reference value can be a previously stored reading or an entered value.
The R&S NRP2 calculates the ratio of measurement result and reference value and
the relative deviation of the measured value from the reference value and indicates the
result.